Product Class
Handheld Pulse Oximeter
  NT1
  NT1A
  NT1B
Handheld Vital Signs Monitors
  NT1D Series
Vital Signs Monitors
  NT2A
  NT2C
Multiparameter Patient Monitors
  NT3A (10.4")
  NT3B (12.1")
  NT3C (10.4")
  NT3E (8.4")
  NT3F (7")
Central Monitoring System
  NTC2700
Monitoring Accessories
  SpO2
  NIBP Cuffs
  ECG Cables
  Temperature Sensors
>> Contact Us
400 Oyster Point Blvd.
Ste. 534
South San Francisco
CA 94080, USA
TEL: +1 650 5883980
FAX: +1 650 5883988
E-mail:
sales@solarismedtech.com
Product -- NT1 (CE Marked, FDA Cleared)
NT1 Handheld Pulse Oximeter
Product Description:

◆Bright, Easy-to-Read Large LED Display
◆Durable, Compact and Lightweight
◆Used for Spot-Check
◆Suitable for Adult and Pediatric

 


Product Features :

SpO2
Measurement Range: 0~100%
Accuracy:±2% during 70%~100%
           0%~69% unspecified


◆Pulse Rate
Measurement range: 30 - 250 bpm
Accuracy: ±1 bpm or 2%
        

◆Visual Displays and Indicators
Sp02 Display:
Red LED, variable brightness
Pulse Rate Display:
Green LED, Variable brightness
Pulse intensity: Red LED 8 Bars
Power Indicator Light: Green
Low Power Voltage: Red and Blink
Low Perfusion Indicator: Red  

◆Audio Indicators
Sensor off (Every 30 seconds)
Pulse Beep
Key-pressing
Low Battery

◆Power Requirements:
Four Standard"AA" Size Batteries
Battery Capacity:≥12 hours

◆Environment:
Operating Temperature: Operation 0~50℃
humidity: ≤95%
Altitude:-390m-5000m
Transport/Storage Temperature:-40℃-70℃
humidity: ≤95%

◆Physical Characteristics:
Dimensions:69mm(W)x130mm(H)x22mm(D)
Net Weight:195g/100g(with batteries/without batteries)
Gross Weight:495g/395g(with batteries/without batteries)

◆Compliance:
1. SpO2: ISO 9919: 2005(E)
2. Safety Standards:
    EN 60601-1: 1990+A1: 1993+A2: 1995+A13: 1996
3. Alarm: IEC60601-1-8: 2005
4. EMC: EN 60601-1-2: 2007, Group 1 Class A
5. Environment : WEEE (2002/96/EC)
6. FDA 510(K)







 
   
Copyright  ©  2010  Solaris Medical Technology, Inc.